Effects of clinical reanalysis in dual energy X-ray absorptiometry reports

Küçük Resim Yok

Tarih

2017

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Baycinar Medical Publ-Baycinar Tibbi Yayincilik

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Objectives: This study aims to assess poor positioning rates of patients during X-ray and the accuracy of the analysis. Patients and methods: In this study, we reanalyzed 323 dual energy X-ray absorptiometry (DXA) reports, by evaluating the scan images for proper patient positioning and scan analysis. We reviewed reports, according to a checklist prepared considering the proposals of Watts and The International Society for Clinical Densitometry official positions for 2013 (which were the same as in 2015). At least two remaining vertebrae were used to derive new bone mineral density and new T-scores. Results: Positioning failures were found in 64.7% of the spine X-rays, 60.5% of the hip X-rays, and 83.9% of X-rays of both regions. A total of 112 (34.7%) spinal DXA images needed new T-score adjustments. T-scores and bone mineral density differed between the first reports and the clinician reanalysis (p<0.001). Conclusion: The error rate in DXA reports was higher than expected. Clinician analysis of DXA reports are important. To obtain a quality DXA report, all healthcare professionals should be trained and reminded about this topic.