Erdogan, OAltun, AOzbay, G2024-06-122024-06-1220010147-8389https://doi.org/10.1046/j.1460-9592.2001.01704.xhttps://hdl.handle.net/20.500.14551/22512This report describes an asymptomatic patient with a DDDR pacemaker system (Inos Cls, Biotronik) that showed intermittent mode reversion to VOO and sometimes to VVI without a detectable source of electromagnetic interference and oversensing. Patients with this Biotronik model should be carefully assessed along with a surface ECG before placing the programming head.en10.1046/j.1460-9592.2001.01704.xinfo:eu-repo/semantics/closedAccessMode ReversionVOODual Chamber PacemakerIntermittent mode reversion to VOO of a dual chamber pacemaker modelArticle241117041705Q2WOS:0001722202000232-s2.0-003516097611816646Q2