Investigation of dose distribution at intersection region of split IMRT fields using two different dosimetric modality

dc.authoridKesen, Nazmiye Donmez/0000-0003-3520-409X
dc.authorwosidUZAL, Mustafa Cem/AAH-2683-2019
dc.authorwosidKesen, Nazmiye Donmez/E-2876-2019
dc.contributor.authorSuncak, Leyla
dc.contributor.authorDonmez Kesen, Nazmiye
dc.contributor.authorUzal, Mustafa Cem
dc.date.accessioned2024-06-12T11:15:33Z
dc.date.available2024-06-12T11:15:33Z
dc.date.issued2015
dc.departmentTrakya Üniversitesien_US
dc.description.abstractOBJECTIVES The purpose of this study is to investigate dose distribution in the overlap region of two split sub-fields generated by TPS. METHODS IMRT treatment fields larger than 15 cm were planned. TPS created two new sub-fields by splitting each original field. IMRT plans of 14 overlapping complementary binary sub-fields were sent to virtual phantom of TPS for quality assurance (QA) calculations by setting all gantry and collimator angles zero. The measurements at linac were taken using 2D-ion chamber lattice and film both placed at 5 cm depth in solid phantom at QA conditions. RESULTS According to different analyze criterias, consistence of dose fluence related to all fields and sub-fields were found respectively between 81%-99.2% and 80.4%-99.7% by 2D-Array ion chamber system. CONCLUSION As a result there was no statistically significant difference between calculated and measured dose distributions in overlap regions of two split fields derived from fields larger than 15 cm.en_US
dc.identifier.doi10.5505/tjoncol.2015.1361
dc.identifier.endpage203en_US
dc.identifier.issn1300-7467
dc.identifier.issue4en_US
dc.identifier.scopus2-s2.0-84955299334en_US
dc.identifier.scopusqualityQ4en_US
dc.identifier.startpage195en_US
dc.identifier.urihttps://doi.org/10.5505/tjoncol.2015.1361
dc.identifier.urihttps://hdl.handle.net/20.500.14551/23979
dc.identifier.volume30en_US
dc.identifier.wosWOS:000379361000003en_US
dc.identifier.wosqualityN/Aen_US
dc.indekslendigikaynakWeb of Scienceen_US
dc.indekslendigikaynakScopusen_US
dc.language.isotren_US
dc.publisherKare Publen_US
dc.relation.ispartofTurk Onkoloji Dergisi-Turkish Journal Of Oncologyen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectSplit Fieldsen_US
dc.subjectGamma-Index Analysisen_US
dc.subjectQuality Assuranceen_US
dc.subjectOverlap Regionen_US
dc.subjectIntensity Modulated Radiation Therapyen_US
dc.titleInvestigation of dose distribution at intersection region of split IMRT fields using two different dosimetric modalityen_US
dc.typeArticleen_US

Dosyalar