In vitro comparison of trueness of 10 intraoral scanners for implant-supported complete-arch fixed dental prostheses
Küçük Resim Yok
Tarih
2020
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Mosby-Elsevier
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
Statement of problem. Digital scanning systems have become popular, but whether these systems are adequate for complete-arch implant-supported fixed dental prostheses is unclear. Purpose. The purpose of this in vitro study was to evaluate the trueness of 10 different dental intraoral scanners. Material and methods. Six implant analogs were installed, and an edentulous mandibular model composed of scannable Type 4 gypsum was scanned with 10 different intraoral scanners (3D Progress, Omnicam, Bluecam, Apollo DI, Planscan, E4D Tech, TRIOS MonoColor Cart, TRIOS Color Cart, TRIOS Color Pod, Lythos), 10 times each after the scan body was placed on the implant abutments. The data obtained were then converted into standard tessellation language format. For the control group, the gypsum model was scanned with an industrial scanner (ATOS Core 80). For trueness, the dental and industrial scanning data packs were analyzed with 3D comparison software. Statistical analyses were performed by using the Kruskal-Wallis and Mann-Whitney U tests. Results. When ranked according to their surface superimposition values, the Color POD, Omnicam, Apollo DI, Color Cart, MonoColor Cart, and Bluecam scanners were found within the range of 31 to 45 mm. This group was followed by E4D, 3D Progress, Lythos, and Planscan, which were found within the range of 82 to 344 mm according to the same criteria. Conclusions. Some of the digital scanners had the necessary performance for the fabrication of complete-arch implant-supported fixed dental prostheses. However, the possibility of data loss producing artifacts should be considered.
Açıklama
Anahtar Kelimeler
Accuracy, Precision
Kaynak
Journal Of Prosthetic Dentistry
WoS Q Değeri
Q2
Scopus Q Değeri
Q1
Cilt
124
Sayı
6