Single lead changed to double lead DDD pacemaker in a patient with diaphragmatic stimulation
Küçük Resim Yok
Tarih
2001
Yazarlar
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
The AA. report on patient who suffered diaphragmatic stimulation due to atrial impulse of a single lead DDD pacemaker system. Although the well-known advantages of single lead DDD systems there are some major drawbacks of this technique to consider before implantation. As it is described in this present case, diaphragmatic stimulation can not be avoided in some individuals. This problem was overcome by inserting a separate atrial lead and programming the pacemaker to DDD mode.
Açıklama
Anahtar Kelimeler
A-V Block; Ddd; Diaphragmatic Stimulation; Pacemaker; Single Lead Vdd, Adult; Angiocardiography; Artificial Heart Pacemaker; Atrioventricular Block; Case Report; Conference Paper; Diaphragm Disease; Dizziness; Dyspnea; Echocardiography; Electrocardiogram; Female; Human