Flame atomic absorption determination of some metal ions in water samples after preconcentration on montmorillonite modified with trioctylamine
Küçük Resim Yok
Tarih
2005
Yazarlar
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Parlar Scientific Publications (P S P)
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
A column method using montmorrilonate with added trioctylamine as adsorbent was developed for the preconcentration and F-AAS determination of Fe, Cd, Zn, Ni, Mn, Cr and Cu ions in wastewater, tap water and river water samples. The influence of aqueous solution pH, amount of montmorillonite, flow rate of elution, different types and least amounts of elutes, and the effects of matrix ions on pre-concentration were investigated. The recoveries were > 95%, and preconcentration factor was 250. The detection limits for Fe, Cd, Zn, Ni, Mn, Cr and Cu were found to be 0.09, 0.19, 0.08, 0.13, 0.21, 0.16 and 0.13 mu g/l, respectively. The relative standard deviation of the determination was < 7%. The adsorption capacities for Fe, Cd, Zn, Ni, Mn, Cr and Cu were calculated to be 12, 11.2, 14.5, 10, 9.7, 13.7 and 14 mg/g, respectively.
Açıklama
Anahtar Kelimeler
Preconcentration, Montmorillonite, Water Samples, Flame Atomic Absorption Spectrometry (F-AAS), Solid-Phase Extraction, Separation, Sorbents, Cadmium, Surface
Kaynak
Fresenius Environmental Bulletin
WoS Q Değeri
Q4
Scopus Q Değeri
N/A
Cilt
14
Sayı
11