Effect of Temperature on the Cyclic Fatigue Resistance and Phase Transformation Behavior of Three Different NiTi Endodontic Instruments

Küçük Resim Yok

Tarih

2024

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Springernature

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Background This study aims to analyze the phase transformation behavior and evaluate the effect of different temperatures on the fracture strength of files. Methodology A total of 108 files were used, and cyclic fatigue tests were performed on HyFlex EDM, VDW.ROTATE, and TruNatomy at three different temperatures (+4 degrees C, +35 degrees C, and +60 degrees C) The time to fracture of files was measured, and the number of cycles was calculated. In addition, the fractured fragment lengths were measured. The Kruskal-Wallis test was used to compare the data that were not normally distributed according to groups. The Mann-Whitney U test was performed for comparisons. P-values <0.05 were considered statistically significant. Results As the temperature increased for all files, the number of cycles to fracture (NCF) decreased. NCF was significantly higher for ROTATE at 4 degrees C and 60 degrees C and HyFlex EDM at 35 degrees C. Conclusions The resistance to cyclic fatigue in all files decreased significantly as the temperature increased.

Açıklama

Anahtar Kelimeler

Nickel Titanium, Martensite, Austenite, Dsc, Cyclic Fatigue, Rotary Nickel-Titanium, In-Vivo, Files

Kaynak

Cureus Journal Of Medical Science

WoS Q Değeri

N/A

Scopus Q Değeri

Cilt

16

Sayı

1

Künye