Importance of Calibration for Effective use of Therapeutic Ultrasound

dc.contributor.authorTuna, Hakan
dc.date.accessioned2024-06-12T11:16:13Z
dc.date.available2024-06-12T11:16:13Z
dc.date.issued2011
dc.departmentTrakya Üniversitesien_US
dc.description.abstractTherapeutic ultrasound is one of the most widely used electrophysical agents. The users are generally not well aware of the details of the technology of ultrasonic generators and cannot be asked to perform the complicated physical measurements to test their equipment. Frequently, the only way to check the performance of the device is the observance of oil or water bubbles on the surface of the piezoelectiric probe. If bubbles are present, the generator is regarded as operating correctly. Many surveys on the output of therapeutic ultrasound equipment have revealed discrepancies between the indicated and actual ultrasonic output of many devices. International safety standards recommend a limit below 30% variation in acoustic intensity for ultrasound therapy equipment. So, further improvements in the accuracy of ultrasound machine calibration are needed. Proper calibration is essential to provide patients with a more accurate ultrasound dosage and, therefore, with a safer and more appropriate treatment. This review highlights the need for users to be aware of the potential inaccuracy of the machine and of the importance of regular calibration. They could be informed better with a systematic training in their education curricula on specific variables of ultrasound machines. Turk J Phys Med Rehab 2011;57:94-100.en_US
dc.identifier.doi10.4274/tftr.57.19
dc.identifier.endpage100en_US
dc.identifier.issn1302-0234
dc.identifier.issue2en_US
dc.identifier.scopus2-s2.0-79959866840en_US
dc.identifier.scopusqualityN/Aen_US
dc.identifier.startpage94en_US
dc.identifier.urihttps://doi.org/10.4274/tftr.57.19
dc.identifier.urihttps://hdl.handle.net/20.500.14551/24238
dc.identifier.volume57en_US
dc.identifier.wosWOS:000296178500008en_US
dc.identifier.wosqualityQ4en_US
dc.indekslendigikaynakWeb of Scienceen_US
dc.indekslendigikaynakScopusen_US
dc.language.isotren_US
dc.publisherGalenos Yayinciliken_US
dc.relation.ispartofTurkiye Fiziksel Tip Ve Rehabilitasyon Dergisi-Turkish Journal Of Physical Medicine And Rehabilitationen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectTherapeutic Ultrasounden_US
dc.subjectCalibrationen_US
dc.subjectPhysical Therapy Agenten_US
dc.subjectEffective Radiating Areaen_US
dc.subjectOutput Poweren_US
dc.subjectTransducersen_US
dc.subjectVariabilityen_US
dc.subjectIntensityen_US
dc.subjectEquipmenten_US
dc.subjectDeviceen_US
dc.subjectTissueen_US
dc.titleImportance of Calibration for Effective use of Therapeutic Ultrasounden_US
dc.typeReview Articleen_US

Dosyalar